What process validation documentation should you request from a manufacturing partner?

A manufacturing partner should be able to share a validation master plan, IQ, OQ and PQ protocols and reports, cleaning validation records, and a change control procedure. Together, these documents show that the process consistently produces products that meet specifications, and that the partner keeps it under control over time.
How much documentation you need depends on what you are outsourcing. A medical device wipe calls for more evidence than a basic cosmetic product, but every category benefits from a clear paper trail. Below we answer the most common questions buyers ask when auditing a manufacturing partner. If you are exploring a new project, you can also take a look at our hygiene products to see what we manufacture.

What do IQ, OQ and PQ reports actually prove?

IQ, OQ and PQ reports prove that equipment is installed correctly, operates within defined limits, and performs consistently under real production conditions. Each stage answers a different question, and together they form the core evidence that a manufacturing process is validated rather than simply assumed to work.

  • Installation Qualification (IQ): confirms the equipment was installed according to specification, with the right utilities, calibration and documentation in place.
  • Operational Qualification (OQ): confirms the equipment works across its full operating range, for example filling volumes, sealing temperatures and line speeds.
  • Performance Qualification (PQ): confirms that the process delivers conforming product repeatedly, usually across several consecutive production batches.

For liquid-based products such as wet wipes, PQ is where the practical questions get answered. Is the liquid dosed evenly into every wipe? Are sachet seals tight? Does the product meet its specification from the first batch to the last?

When you review these reports, look for pre-approved acceptance criteria. Good protocols define what “pass” means before the testing begins, not afterwards. Ask also which batch sizes and which product variants were covered, because a validation that reflects your real production conditions is worth far more than a generic one.

Which validation documents should a manufacturing partner be able to share?

A reliable manufacturing partner should be able to share a validation master plan, IQ, OQ and PQ protocols and reports, cleaning validation records, packaging and sealing validation, and a change control procedure. Some details may be confidential, but the existence and structure of these documents should never be a mystery.

Use this checklist when requesting documents:

  1. Validation master plan: the overview of what is validated, how, by whom and on what schedule.
  2. Process validation protocols and reports: the IQ, OQ and PQ evidence for the equipment and process used for your product.
  3. Cleaning validation: evidence that residues and microbial contamination are controlled between products, which matters for any partner running several formulas on shared lines.
  4. Packaging and seal validation: proof that sachets, bottles and tubes protect the product through its shelf life.
  5. Equipment calibration and maintenance records: showing that critical instruments stay accurate over time.
  6. Change control and deviation procedures: showing how unplanned events and planned modifications are handled.
  7. Relevant certificates and audit reports: such as ISO 9001, ISO 13485 and ISO 22716 certificates.

It is also worth asking for a sample of a completed validation report, with sensitive customer data removed. Seeing the real format tells you more than any verbal assurance. In our own work with healthcare products and medical devices, we handle pilot production and validation as part of the development process, so the documentation grows together with the product.

How can you tell if validation documentation is solid?

Solid validation documentation is traceable, pre-approved, data-backed and specific to your product. It links every claim to recorded evidence, shows who approved what and when, and includes clear acceptance criteria. Vague summaries, undated reports and missing signatures are signs that the validation may not hold up under scrutiny.

A quick quality check can be built around these questions:

  • Was the protocol approved before the validation runs were performed?
  • Are the raw data, batch records and test results attached or referenced?
  • Were any deviations recorded, investigated and closed with a documented rationale?
  • Does the report state a clear conclusion against the acceptance criteria?
  • Is the document under version control, with named approvers from quality and production?

Perfect-looking reports with no deviations at all can actually be a warning sign. Real validation work occasionally surfaces small problems, and a mature quality system documents how they were resolved. What you want to see is honesty and control, not flawlessness.

A well-run partner audit is the best way to verify all of this. Walk the production floor, compare the documents with what you see, and talk to the people who run the process daily. If you would like to see how we handle this, you are welcome to contact us and arrange a discussion with our experts.

What validation evidence do ISO 13485 and MDR require?

ISO 13485 requires that processes whose output cannot be fully verified by later inspection be validated, with documented procedures, records and defined criteria. The EU Medical Device Regulation (MDR) builds on this by requiring that validation evidence be part of the technical documentation and quality management system that supports your device.

In practice, this means a medical device manufacturer should be able to show:

  • Which processes were identified as requiring validation, and why
  • Approved protocols, acceptance criteria and results for each validated process
  • Qualification of equipment and, where relevant, software used in production
  • Records of personnel training and approved procedures for running the process
  • Links between validation results and the risk management file, in line with ISO 14971

Sterile or microbiologically sensitive products, sealing processes and filling steps are typical candidates for formal validation. The MDR also expects documented control of suppliers and subcontracted processes, so your contract manufacturer’s validation records become part of your own regulatory story.

For cosmetics, ISO 22716 sets good manufacturing practice expectations instead. It does not use the same validation vocabulary, but it still expects controlled, documented and repeatable production, including cleaning, equipment control and change management. Our processes are covered by ISO 13485 and ISO 22716, among other standards, and we support our customers with the product documentation these frameworks require.

When does a validated process need revalidation?

A validated process needs revalidation when something changes that could affect product quality, such as new equipment, a modified formula, different packaging materials, a new site, or a shift in process parameters. Revalidation may also be required after repeated deviations or trend data showing the process is drifting.

Typical triggers include:

  1. Changes to the formulation, raw material suppliers or specifications
  2. Equipment replacement, major repairs or relocation
  3. Changes to packaging materials, sachet films or closure systems
  4. Adjustments to critical process parameters such as dosing, sealing or mixing
  5. Transfer of production to a new line or facility
  6. Negative quality trends, complaints or recurring deviations

This is where change control becomes essential. A good partner assesses every proposed change for its quality impact and decides whether it needs no action, partial requalification or full revalidation. Ask how they document that decision, and who must approve it. If you are transferring an existing product to a new manufacturer, expect a fresh validation at the new site, even if the product itself is unchanged.

Many partners also run periodic reviews, for example annually, to confirm that validated processes remain in a state of control. These reviews give you confidence that validation is a living practice, not a one-time project.

What red flags suggest weak validation practices?

The clearest red flags are missing or unsigned protocols, validation based only on a single batch, no change control procedure, and a reluctance to share documentation. A partner who cannot explain how their processes were validated is probably relying on habit rather than evidence.

Watch for these warning signs during evaluation and audits:

  • Retrospective paperwork: protocols written or signed after the validation runs took place.
  • No acceptance criteria: reports that describe results but never state what counted as a pass.
  • Generic documents: validation that covers “a similar product” rather than yours.
  • Weak cleaning controls: no cleaning validation on shared lines, or no records of changeover checks.
  • Informal change handling: process changes made without documented impact assessment.
  • Outdated records: calibration overdue, or validation reports that have never been reviewed since the first approval.
  • Evasive answers: hesitation to show the validation master plan or sample reports.

One red flag does not always mean a bad partner, but a pattern should make you pause. The strongest partners welcome these questions because they have nothing to hide and a lot to show.

If you are comparing manufacturing partners for wet wipes or other liquid-based products, we are happy to walk you through our validation approach and documentation. Send us a contact request, and one of our experts will get back to you. You can also learn more about our hygiene products to see how our quality-driven production supports your brand.